Mouse Coagulation Factor XII enzyme-linked immunoassay kit

CAT: EM0041 Datasheet
Specification 96 Test
Sensitivity 2.35 pg/ml (10 μl)
Standard Curve Range 13.72~10000 pg/ml
Standard Curve Gradient 7 Points/3 Folds
Number of Incubations 2
Sample Volume 10 μl
Type Fully Ready-to-Use
Operation Duration 120min
pg/ml O.D. Average Corrected
0.00 0.0164 0.0142 0.0153
13.72 0.0253 0.0224 0.0239 0.0086
41.15 0.0286 0.0276 0.0281 0.0128
123.46 0.0472 0.0489 0.0481 0.0328
370.37 0.1131 0.1289 0.1210 0.1057
1111.11 0.3938 0.4223 0.4081 0.3928
3333.33 1.3840 1.4040 1.3940 1.3787
10000.00 3.3980 3.3740 3.3860 3.3707

Precision

Intra-assay Precision Inter-assay Precision
Sample Number S1 S2 S3 S1 S2 S3
22 22 22 6 6 6
Average(pg/ml) 186.2 1021.9 3453.9 143.3 846.7 2856.7
Standard Deviation 5.7 43.2 230.7 6.7 56.7 113.3
Coefficient of Variation(%) 3.0 4.2 6.7 5.5 6.6 4.0

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.

Spike Recovery

The spike recovery was evaluated by spiking 3 levels of Mouse Coagulation Factor XII into health Mouse serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 93% to 106% with an overall mean recovery of 100%.

Sample Values

Sample Matrix Sample Evaluated Range (μg/ml) Detectable (%) Mean of Detectable (μg/ml)
Serum 30 13.00-60.31 100 33.76

Serum/Plasma – Thirty samples from apparently healthy mice were evaluated for the presence of Coagulation Factor XII in this assay. No medical histories were available for the donors. n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.

Background: Coagulation Factor XII

Coagulation factor XII (F12) is a serum glycoprotein that plays multiple roles in initiating blood coagulation, fibrinolysis, and the production of bradykinin and angiotensin.