Human CD63 ELISA Kit检测试剂盒
规格 | 96 Test |
---|---|
灵敏度 | 0.24 pg/ml (50 μl);6.71 pg/ml (10 μl); |
标准曲线范围 | 6.86~5000 pg/ml |
标准曲线梯度 | 7点/3倍 |
孵育次数 | 2 |
可检测样本 | 可溶性物质的液相样品。例如:血清,血浆,细胞培养上清,组织研磨液等。 |
加样体积 | 50 μl/10 μl |
类型 | 完全即用型 |
操作时长 | 120min |

pg/ml | O.D. | Average | Corrected | |
---|---|---|---|---|
0.00 | 0.0151 | 0.0128 | 0.0140 | |
6.86 | 0.0286 | 0.0280 | 0.0283 | 0.0144 |
20.58 | 0.0572 | 0.0582 | 0.0577 | 0.0438 |
61.73 | 0.1475 | 0.1458 | 0.1467 | 0.1327 |
185.19 | 0.4194 | 0.3984 | 0.4089 | 0.3950 |
555.56 | 1.2120 | 1.1120 | 1.1620 | 1.1481 |
1666.67 | 2.6950 | 2.5740 | 2.6345 | 2.6206 |
5000.00 | 3.8760 | 3.8910 | 3.8835 | 3.8696 |
精度
Intra-assay Precision | Inter-assay Precision | |||||
Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
22 | 22 | 22 | 6 | 6 | 6 | |
Average(pg/ml) | 336.6 | 1807.6 | 5114.7 | 325.3 | 1844.1 | 5419.5 |
Standard Deviation | 10.2 | 76.5 | 268.5 | 11.8 | 98.5 | 124.0 |
Coefficient of Variation(%) | 3.0 | 4.2 | 5.2 | 3.6 | 5.3 | 2.3 |
板内精密度:3 个已知浓度的样本酶标板内重复测定 20 次,评估酶标板内的精密度。
板间精密度:3 个已知浓度的样本酶标板间重复检测 6 次,评估酶标板间的精密度。
加标回收
健康人血清加入 3 个不同浓度水平的人 CD63,未加人 CD63 的血清作为本底,计算回收率。回收率的范围从 93 % 至 119 %,平均回收率为 105 %。
样本值
Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
---|---|---|---|---|
Serum | 30 | 570.74-1262.25 | 100 | 787.53 |
应用本试剂盒,检测 30 个健康志愿者的血清样本,志愿者的用药史不详。n.d. = 测不到浓度值。样本的浓度值低于灵敏度被认为测不到浓度值。
产品数据表
背景: CD63
CD63,也称为 LAMP-3 或 ME491(黑色素瘤相关抗原),是蛋白质运输蛋白跨膜蛋白超家族的 30-60 kDa 成员。CD63 普遍表达并存在于晚期内吞囊泡中,但细胞活化后也存在于质膜上。CD63 与其他膜蛋白或接头蛋白的相互作用调节细胞活性,如粘附、迁移和脱颗粒。人 CD63 的细胞外区域与小鼠和大鼠 CD63 分别具有 67% 和 65% 的氨基酸序列同一性。