Human IDO1 ELISA Kit检测试剂盒
规格 | 96 Test |
---|---|
灵敏度 | 1.61 pg/ml (50 μl);3.04 pg/ml (10 μl) |
标准曲线范围 | 6.86~5000 pg/ml |
标准曲线梯度 | 7点/3倍 |
孵育次数 | 2 |
加样体积 | 50 μl/10 μl |
类型 | 完全即用型 |
操作时长 | 120min |

pg/ml | O.D. | Average | Corrected | |
---|---|---|---|---|
0.00 | 0.0180 | 0.0162 | 0.0171 | |
6.86 | 0.0263 | 0.0268 | 0.0266 | 0.0095 |
20.58 | 0.0455 | 0.0428 | 0.0442 | 0.0271 |
61.73 | 0.0997 | 0.0941 | 0.0969 | 0.0798 |
185.19 | 0.2684 | 0.2517 | 0.2601 | 0.2430 |
555.56 | 0.7869 | 0.7186 | 0.7528 | 0.7357 |
1666.67 | 1.9390 | 1.8720 | 1.9055 | 1.8884 |
5000.00 | 3.6660 | 3.7050 | 3.6855 | 3.6684 |
精度
Intra-assay Precision | Inter-assay Precision | |||||
Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
22 | 22 | 22 | 6 | 6 | 6 | |
Average(pg/ml) | 119.6 | 549.6 | 1673.2 | 108.1 | 542.6 | 1597.0 |
Standard Deviation | 5.6 | 33.6 | 90.3 | 2.9 | 22.2 | 63.4 |
Coefficient of Variation(%) | 4.7 | 6.1 | 5.4 | 2.7 | 4.1 | 4.0 |
板内精密度:3 个已知浓度的样本酶标板内重复测定 20 次,评估酶标板内的精密度。
板间精密度:3 个已知浓度的样本酶标板间重复检测 6 次,评估酶标板间的精密度。
加标回收
健康人血清加入 3 个不同浓度水平的人 IDO1,未加人 IDO1 的血清作为本底,计算回收率。回收率的范围从 81 % 至 103 %,平均回收率为 92 %。
样本值
Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
---|---|---|---|---|
Serum | 30 | 62.75-652.09 | 100 | 291.93 |
应用本试剂盒,检测 30 个健康志愿者的血清样本,志愿者的用药史不详。n.d. = 测不到浓度值。样本的浓度值低于灵敏度被认为测不到浓度值。
产品数据表
背景: IDO1
吲哚胺 2,3-双加氧酶 (IDO) 是一种含血红素的细胞内双加氧酶,催化必需氨基酸 L-色氨酸降解为 N-甲酰基犬尿氨酸。这种降解是 L-犬尿氨酸通路的第一步也是限速步骤。IDO 在树突状细胞、巨噬细胞、小胶质细胞、嗜酸性粒细胞、成纤维细胞、内皮细胞和大多数肿瘤细胞中广泛表达。在免疫细胞中,其表达主要由 IFN-γ、IFN-α、IFN-β 和 IL-10 等细胞因子诱导。IDO 具有抗菌功能,因为它会降低炎症环境中必需氨基酸色氨酸的可用性。最近的研究表明,IDO 在感染、妊娠、移植、自身免疫和肿瘤形成过程中诱导免疫抑制。