Human CCL13/MCP-4 ELISA Kit检测试剂盒
| 规格 | 96 Test |
|---|---|
| 灵敏度 | 0.03 pg/ml (50 μl);0.18 pg/ml (10 μl); |
| 标准曲线范围 | 0.69~500 pg/ml |
| 标准曲线梯度 | 7点/3倍 |
| 孵育次数 | 2 |
| 可检测样本 | 可溶性物质的液相样品。例如:血清,血浆,细胞培养上清,组织研磨液等。 |
| 加样体积 | 50 μl/10 μl |
| 类型 | 完全即用型 |
| 操作时长 | 120min |
| pg/ml | O.D. | Average | Corrected | |
|---|---|---|---|---|
| 0.00 | 0.0085 | 0.0090 | 0.0088 | |
| 0.69 | 0.0257 | 0.0247 | 0.0252 | 0.0165 |
| 2.06 | 0.0596 | 0.0624 | 0.0610 | 0.0523 |
| 6.17 | 0.1737 | 0.1756 | 0.1747 | 0.1659 |
| 18.52 | 0.5100 | 0.4768 | 0.4934 | 0.4847 |
| 55.56 | 1.3790 | 1.2690 | 1.3240 | 1.3153 |
| 166.67 | 3.0120 | 2.9570 | 2.9845 | 2.9758 |
| 500.00 | 4.2130 | 4.3809 | 4.2970 | 4.2882 |
精度
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average(pg/ml) | 11.6 | 60.1 | 194.9 | 10.7 | 53.7 | 177.4 |
| Standard Deviation | 0.4 | 1.9 | 6.6 | 0.6 | 2.3 | 9.1 |
| Coefficient of Variation(%) | 3.1 | 3.2 | 3.4 | 5.2 | 4.3 | 5.2 |
板内精密度:3 个已知浓度的样本酶标板内重复测定 20 次,评估酶标板内的精密度。
板间精密度:3 个已知浓度的样本酶标板间重复检测 6 次,评估酶标板间的精密度。
加标回收
健康人血清加入 3 个不同浓度水平的人 CCL13/MCP-4,未加人 CCL13/MCP-4 的血清作为本底,计算回收率。回收率的范围从 80 % 至 120 %,平均回收率为 100 %。
样本值
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Serum | 30 | 22.30-140.00 | 100 | 75.86 |
应用本试剂盒,检测 30 个健康志愿者的血清样本,志愿者的用药史不详。n.d. = 测不到浓度值。样本的浓度值低于灵敏度被认为测不到浓度值。
产品数据表
背景: CCL13/MCP-4
人 MCP-4,也称为 CCL13,编码 98 个氨基酸 (aa) 的前体蛋白,具有 23 个氨基酸的疏水信号肽,该信号肽被切割后产生 8 kDa、75 个氨基酸的成熟 MCP-4。成熟的 MCP-4 缺乏任何潜在的 N-糖基化位点,并与其他人类 MCP 蛋白共享焦谷氨酸脯氨酸基序。